New Features of Rooster Monitor100 Benefit Critical Containment Applications
- Event data logging with USB export
- Configurable alarm tones
- User's preferences can be uploaded/downloaded via USB
Rooster™ Airflow Alarm & Monitor Platform - V1.04 New Feature Rollout
Data and event logging addresses protocols where users need to show that they’re in compliance. The event log can be viewed on-screen and also exported to a USB drive in standard file format for compliance record keeping.
The ability to modify alarm tones, per alarm, gives users more information instantly and makes alarming tolerable in select environments where loud sounds cause discomfort.
The Rooster™ configuration file can be downloaded to a USB thumb drive for future use and also uploaded to other Monitor100s in your facility. The capacity to save and recall chosen settings saves valuable setup time and ensures operational consistency.
As current users know, you can firmware upgrade your Rooster™ via USB to stay up to date and take advantage of these new feature benefits.
Note: A Rooster™ with firmware preceding version 1.04 may be upgraded to version 1.04, however, existing calibration data of the chemical fume hood will be lost, and all settings will be reset to factory defaults. As a result, this update should be done during annual certification, where the certifier would be recalibrating anyway, OR, if the client approves forcing a recalibration.